Operation
End of process
Confirm or cancel all changes with
Process parameters
Process parameters
Tap the
The process parameters menu opens.
Touch the function button of the desired parameter and adjust the value using the
on-screen keyboard, see also "User input" on page 95.
Parameters
Heating tape
Heating tape
automatic
Evacuation
after draining
collector
Educt empty:
additional
Evaporation time
Forced draining
collector
Standby timer
after educt
container runs
empty
Specify here whether the distillate and/or residue are
drained at the end of the process. Touch the corresponding
check box to enable (check mark) or disable (blank) the
option.
Accept
button in the footer.
Function
Performance of the heating tape during operation.
Range: 10 – 100% in increments of ten.
Automatically switch on heating tape yes/no.
Range: on / off.
Evacuation period of the bypass line after draining the
collector vessel.
Range: 1 – 9999 s.
When the educt container runs empty: Additional evapo-
ration time for evaporation of the remaining educt in the
evaporation flask.
Range: 1 – 9999 s.
Forced draining collector active: Collector will be drained
after 1 – 9999 seconds, regardless of the filling level.
Range: 1 – 9999 s.
Activate/deactivate forced draining of the collector (deter-
mines previous parameter!).
Range: active / inactive
Time until the standby mode is activated after the educt
container runs empty.
Range: 1 – 9999 s.
Parameters Hei-VAP
page 122.
Datalogging
, see section "Data Recording" on page 108.
Save
, see section "Recipes" on page 123.
Cancel
or
.
, see section "Parameters Hei-VAP" on
129