Commissioning
WARNING: Risk Of Injury, Damage To Property
Due to faulty components, faulty assembly of components or faulty
connections, there is a risk that pressure-applied components will
suddenly loosen or rupture during an overpressure test!
Start test
Touch the
Confirm the subsequent confirmation prompt with
test:
During the test, the system pressure is built up to the defined maximum
Settings
pressure (
maximum of 30 seconds!
When the maximum pressure is reached, a measurement is made for a period
of 240 seconds to determine whether the system pressure drops below a
defined limit value.
After this time has elapsed, the system message
successfully completed
required values have been achieved.
If the defined maximum pressure is not reached during the overpressure test
or the system pressure falls below a defined limit value, the error message
Overpressure test failed
the display.
The test can be stopped at any time by
Press
to close the menu.
Before performing an overpressure test, ensure that all compo-
nents are properly installed.
In particular, make sure that all screw connections are properly
tightened.
Before starting the overpressure test, close all safety covers,
hoods and doors of the entire system.
Ensure that third parties are clear of the danger zone.
Use the required personal protective equipment (eye protection,
hand protection, etc.).
function button.
menu). This pressure level must be reached within a
is shown on the display, provided that all
appears. In this case, follow the instructions on
Stop test
to carry out the overpressure
Overpressure test
.