Table 1. Operating values of the Model B35200 Neurostimulator
Parameter
Amplitude
Pulse width
Rate
Cycling
BrainSense operating values:
Sensing noise floor
Input sensing range
a
Interlocks will prevent the use of some parameter combinations.
b
Refer to the appropriate clinician application programming guide for additional rate considerations.
Sensing noise floor performance assumes sensing with electrodes adjacent to stimulation, delivered
c
through a monopolar configuration. For example, stimulate 1-C+ and sense 0-2 with stimulation
parameters <5 mA (at 1000 Ω impedance) and <120 µs.
8 English  B35200 2020-01-01
c
Operating range
0 to 25.5 mA
20 to 450 µs
b
2 to 250 Hz
Off, On: 0.1 s to 24 hr
<300 nV/rtHz
0.55 to 400 µVrms
a