3 Alignment and Adjustment

3.1 Connecting the Tube Adapter

1 Mount the foot to the tube adapter (lightly tighten the screws).
2 Plug the tube adapter plug into the knee joint (only one polarity possible).
3 Push the protruding cable into the tube and push the tube adapter into the C-Leg compact. In doing so, take care
to achieve the minimum insertion depth (see following table).
max. 75 kg
Error during alignment of the prosthesis. Errors during the alignment of the prosthesis can lead to mal-
CAUTION
function of the joint up to loss of function due to structural failure. This can cause the patient to fall. With the 3C96-1
and 3C86-1 models, the scaling must be positioned in front of the tube. The 2R80=110 Tube Adapter is approved for a
maximum weight of 100 kg / 220 lbs.
4 Turn the foot slightly outward (adjustment range ±30° to suit the needs of the patient) and pre-fasten the screws
lightly at the tube clamp.
C-Leg compact Prosthetic System
1
50 mm
max. adjust-
ment range
2
75 kg – 100 kg
45 mm
max. adjust-
ment range
3
100 kg – 125 kg
40 mm
max. adjust-
ment range