MANUAL
Pressure comparison pumps
Avoid exposure to the following factors:
• Direct sunlight or proximity to hot objects
• Mechanical vibra�on, mechanical shock (pu�ng it down hard)
• Soot, vapour, dust and corrosive gases
• Poten�ally explosive environments, flammable atmospheres
• Corrosive liquids
Store the model
LR- Cal
a loca�on that fulfils the condi�ons listed above. If the original packaging is not available, pack and
store the instrument as described below:
1. Wrap the instrument in an an�sta�c plas�c film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If shored for a prolonged period of �me (more than 30 days), place a bag, containing a
desiccant, inside the packaging.
6. Commissioning, opera�on
6.1 Unpacking the hydraulic pressure comparison test pump
As soon as possible a�er delivery open the packaging of the hydraulic pressure comparison test
pump and check that you have all the items detailled in the packing list (see chapter 4.2 „Scope of
delivery"). As you are unpacking the items, examine them for signs of damage or breaking during
transit.
If any items are missing get in touch immediately with DRUCK & TEMPERATUR Leitenberger GmbH
to inform us of the shortage.

6.2 Environmental requirement

When si�ng the hydraulic pressure comparison test pump not in a temperature controlled
laboratory, look for an area that sa�sfies the following criteria as much as possible:
• A constant temperature area free from draughts and sources of heat or cold
• An area free from noise and vibra�on, constantly used pathways
• A clean dry area free from corrosive liquids or vapours
A strong, stable, level table or workbench with the capability of suppor�ng the system with
sufficient space to operate is required.
6.3 Decision for an opera�ng medium
You may use opera�ng
If the pump is used with mineral oil (par�cularly oil with high viscosity), the sinter
filter in the reservoir base has to be dismounted:
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
LSP-H
hydraulic pressure comparison test pump in its original packaging in
liquids
only, which are described in chapter 3.2.
DRUCK
&
TEMPERATUR