The higher the temperature difference between the sensor and the liquid being tested,
the longer stabilisation of the measured value will take. Stabilisation can take between
ten seconds and five minutes.
8. To retain the measured value in the display when the sensor is removed from the liquid,
press the
button (5). The Data Hold Symbol "HOLD" (K) appears in the LCD display.
9. Press the
button again to deactivate the data hold function and the measured value is
deleted. The Data Hold Symbol "HOLD" (K) will disappear from the LCD display (6).
The automatic temperature compensation (ATC) always ensures precise measure-
ments even where there are differing liquid temperatures.
Displaying maximum/minimum values and switching temperature units
1
1. The meter can display the maximum and minimum values of a measurement.
2. Press and hold the
LCD display.
3. Press the
button briefly to switch between the maximum and minimum values.
4. Press and hold the
5. Press and hold the
g) Advanced Settings
1. Press the
2. Then press and hold the
ton (5), to select "SaLC" (salinity compensation) or AltC (altitude compensation). Press the
button (4) to confirm your selection.
button (5) or the
3. Press the
tude compensation. The salinity compensation can be set in a range between 0 and 50 ppt.
The altitude compensation is adjustable between the values of 0 and 20,000 ft.
4. Confirm the value set by pressing the
button until "MAX" (C) and "MIN" (A) flash simultaneously in the
button (4) to exit display of the MIN/MAX values.
button (10) to toggle the temperature units between °C and °F.
button (10), to select operating mode "DO".
button (4) to call up the selection menu. Press the
button (10), to set the degree of the salinity or the alti-
button (4).
but-