5.1. Specifications
Infrared-Range
Range
Distance to Spot
size
Resolution
Accuracy
Note!
Accuracy at 18 ° C to 28 ° C and Humidity less than 80%.
Field of View:
Make sure that the target to be measured is bigger than the
laser beam. The smaller the target, the closer you should be
located to it during measurement. If the accuracy is not given,
make sure that the target is 2 x larger than the laser beam.
Type-K Range
Display
symbol
Measurement
range
Resolution
Accuracy
-50 ... + 2200° C
50 : 1
0.1° C < 1000° C ; 1° C > 1000° C
Tolerance
+/- 1% + 1° C
+/- 1.5%
+/- 3.5%
TK
-50 ... + 1370° C
0.1° C < 1000° C ; 1° C > 1000° C
Tolerance
+/- 0.5% + 1.5° C
+/- 2.5° C
-58-
Range:
20 ... 500° C
500 ... 1000° C
1000 ... 2200° C
Range:
< 1000° C
> 1000° C

Werbung