Error messages/operating state (LED display)

The relevant device status and error overview is clearly displayed in the Sta-
tus overview window. Open the window via the "Status" button in the Mea-
sured value display window.
The meaning of the displays and possible corrective measures for error mes-
sages are listed in the online help (use F1 to call the Help). Further helpful tips
and setting information can be found in the online help in the FAQs section.
Depending on the display mode, various error messages can be displayed
instead of the measured value in the LCD-display of the MP85A process con-
troller (or the PME Assistant).
Current errors are displayed continuously. Press the
ROR" display mode.
Error message
TransdErr
ADC Ovfl.
MP85A(DP) FASTpress, MP85A(DP)−S EASYswitch
Cause
Input signal overflow
Transducer not connected
Incorrect transducer connection
Amplifier does not match transducer
type
No sense leads connected
Input signal of measurement chan-
nel A/D converter too large
key to go to the "ER-
+
Remedy
Connect transducer
See pin assignment
Connect sense leads
Adapt hardware measuring range
(menu: Preparing for measure-
ment/Amplifier/Transducer)