Herunterladen Inhalt Inhalt Diese Seite drucken

Operation; Maintenance - Mettler Toledo InPro 6000/InTap Serie Kurzanleitung

Optische o2-sensoren
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
InPro 6000 / InTap Optical O
For M400 2-w, M400 FF and M400 PA, the optical
oxygen sensor needs to be powered separately with
an appropriate power supply.
a
Warning: Never connect an InPro 6860 i with mA
output to a nA input. This could damage the electron-
ics of the sensor and / or the instrument the sensor is
connected to.
The RS-485 signal also includes a MODBUS RTU sig-
nal, for a MODBUS installation and additional material
including configuration tools can be downloaded at
www.mt.com/pro-ODO-documentation
4

Operation

The sensor is supplied ready for use. Before using
the senor, remove the protecting cap. For continuous
applications we recommend periodic recalibration in
line with your requirements on accuracy.
In case of any failure or malfunction, the equipment
in which the sensor is installed, needs to be in a safe
condition (depressurized, empty, rinse, vent, etc.).
5

Maintenance

The incorporated Intelligent Sensor Management (ISM)
monitors and displays the remaining lifetime of the
sensing element as the Dynamic Lifetime Indicator.
Depending on the aggressiveness (CIP / SIP) of the
application conditions the sensor's sensing element,
the OptoCap experiences a degree of stress. Typically,
after 6 to 12 months this part needs to be exchanged
followed by a zero point and air calibration.
This calibration can be performed by using the trans-
mitter's calibration menu and / or using METTLER
TOLEDO's calibration and maintenance software tool,
iSense.
www.mt.com/iSense
www.mt.com/pro-transmitters
InPro 6000 / InTap Optical O
© 06 / 2019 Mettler-Toledo GmbH.
Sensors
2
Sensors
2
23
30 313 394 B
Printed in Switzerland
en

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Inpro 6870iInpro 6960iInpro 6860i

Inhaltsverzeichnis