4�1�5 Modbus Installation
The InSUS H60i offers a Modbus RTU connection that is
identical to METTLER TOLEDO's InPro 6860i optical oxygen
sensor. For correct installation, please follow the instruc-
tions of the control system. For programming. please follow
the information given in the "Modbus Programmers' Guide
InPro 6860i" which is available at your local METTLER
TOLEDO Partner.
The Modbus interface allows using the full functionality of
the sensor head including all Intelligent Sensor Manage-
ment (ISM™) functions.
5

Operation

5�1

Configuration

5�1�1 Sensor Head Detection
Before installing an optical
manual for the transmitter and configure the transmitter for
automatic sensor detection. Date and time must be set
correctly in the
such as DLI / ACT will not work properly.
If the
setup might be corrupted.
h
Note: Each time the sensor head is connected with a new
single-use oxygen sensor and diagnostic functions are
activated, DLI / ACT settings must be reset on the transmit-
ter. Please refer to the manual of the transmitter.
5�1�2 Sampling Rate
Optical oxygen sensors do not measure permanently. Each
measurement cycle has a duration of approx. 1 second.
The measurement interval can be set to any value between
1 and 60 seconds. Please choose the appropriate setting.
The default setting is 3 seconds,
applications.
5�1�3 LED Mode
The measurement can be switched off if the system is not
needed.
When the sensor head is not measuring, the
LED is off. In this state the sensor head sends a constant
measurement value of −1 % air to the transmitter and the
transmitter is set to the "Hold mode". To configure the "Hold
mode" please refer to the transmitter manual.
Automatic switch off at high temperature
If the LED mode is set to "Auto" (default setting) the
head
LED will be switched off as soon as a specific process
temperature is reached.
Temperatures
InSUS H60i
InSUS H60i
30 802 188 A
transmitter, otherwise diagnostic functions
wrong date and time are set, the calibration and
Maximum operating
temperature
60 °C (140 °F)
sensor
head, please refer to the
which is sufficient for most
Default switch off
temperature
60 °C (140 °F)
© 09 / 2022 METTLER TOLEDO
Printed in Switzerland
InSUS H60i
sensor head
sensor