Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic Itrel 4 Implantationsanleitung Seite 8

Programmierbarer neurostimulator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Filename Date Time
UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Table 1. Operating values for the Itrel 4 Model 37703 neurostimulator
Programmable parameter
Number of programs
Electrode configuration
Amplitude
Amplitude – upper patient limit
Amplitude – lower patient limit
Pulse width
Pulse width – upper patient limit
Pulse width – lower patient limit
Rate
Rate – upper patient limit
Rate – lower patient limit
SoftStart/Stop
Cycling
Interlocks will prevent the use of some parameter combinations.
a
6 English
37703 2012-03
2012-03
Template version: 12-13-2011
Operating range and resolution
1
1 to 4 electrodes as anode (+), cathode (-), or Off; case
as anode or Off
0 to 10.5 V with 0.05 V or 0.1 V resolution
Tracking limit: Programmed value +0 to +4.0 V (0.5 V
resolution)
Custom limit: Programmed value to 10.5 V (same
resolution as amplitude)
Custom limit: 0 V to the programmed value (same
resolution as amplitude)
60 to 450 µs (10 µs resolution)
Tracking limit: Programmed value +0 to +150 µs (30
µs resolution)
Custom limit: Programmed value to 450 µs (10
µs resolution)
Custom limit: 60 µs to the programmed value (10
µs resolution)
2 to 130 Hz (resolution: 1 Hz from 2 Hz to 10 Hz, 5 Hz
from 10 Hz to 130 Hz)
Tracking limit: Programmed value +0 to +50 Hz (10
Hz resolution)
Custom limit: Programmed value to 130 Hz (same
resolution as rate)
Custom limit: 2 Hz to the programmed value (same
resolution as rate)
Off, On: 1-, 2-, 4-, or 8-second ramp duration
Off, On: 0.1 s to 30 min (resolution: 0.1 s from 0.1 s to 1 s,
1 s from 1 s to 1 min, 1 min from 1 min to 30 min)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
a
M948681A002

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis