Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic Itrel 4 Implantationsanleitung Seite 12

Programmierbarer neurostimulator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Filename Date Time
UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
connector pins may damage the setscrews and the extension connector pins
will not be seated fully into the connector block.
Limit counter-clockwise rotations of the neurostimulator setscrews when
retracting them. Too many counter-clockwise rotations may disengage the
setscrew from the connector block.
Figure 1. Insert the extension connector pins fully into the neurostimulator.
4. Fully insert the torque wrench into each self-sealing grommet of the connector block and
tighten each setscrew (Figure 2).
Note: The torque wrench must be oriented to the same angle as the setscrew
(Figure 2).
#
Cautions:
Be sure the torque wrench is fully inserted into the self-sealing grommet. If the
torque wrench is not fully inserted, the setscrew may be damaged, resulting in
intermittent or loss of stimulation.
Before tightening the setscrews, ensure that the extension connector pins are
inserted into the connector block to prevent damaging the connector block.
Verify that each leaf of the self-sealing grommet is closed after the torque
wrench is withdrawn. If fluid leaks through a grommet seal that is not fully
closed, the patient may experience shocking, burning, or irritation at the
neurostimulator implant location, or intermittent stimulation or loss of
stimulation may occur.
10 English
37703 2012-03
2012-03
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Template version: 12-13-2011
M948681A002

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis