Herunterladen Diese Seite drucken

Storz TP 100 Gebrauchsanweisung Seite 77

Werbung

71
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
The TELE PACK X LED model TP 100 is designed for operation in the electromagnetic environment defined below.
The user of the device should make sure that it is used in such an environment.
Conducted RF
3 V
disturbances acc. to
150 kHz to < 80 MHz
IEC 61000-4-6
3 V/m
Radiated RF disturbances
acc. to IEC 61000-4-3
80 MHz to 2.5 GHz
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects and
people.
Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF
compliance level above, the device should be observed to verify normal operation. If unusual performance characteristics are observed, additional measures
may be necessary, such as a modified alignment or a different location for the device.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Information on
electromagnetic
rms
rms
Portable and mobile RF communications equipment should be used no
closer to any part of the TELE PACK X LED model TP 100, including
cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distances:
B
150 kHz to < 80 MHz
B
80 MHz to < 800 MHz
E
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
According to a study, the field strength of stationary transmitters at all
frequencies on site
a
should be less than the conformity level
Interferences may occur in the vicinity of equipment marked with the
following symbol:
7
b
.

Werbung