Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
The TELE PACK X LED model TP 100 is designed for operation in the electromagnetic environment defined below.
The user of the TELE PACK X LED model TP 100 should make sure that it is used in such an environment.
Electrostatic discharge (ESD)
± 4 kV contact discharge
acc. to IEC 61000-4-2
± 8 kV air discharge
Electrical fast transient/burst
± 2 kV for power supply lines
IEC 61000-4-4
± 1 kV for input/output lines
± 1 kV outer conductor –
Surges
outer conductor
acc. to IEC 61000-4-5
± 2 kV outer conductor –
earth
Voltage dips, short-time
interruptions, and fluctuations
<5 % U
of the supply voltage acc. to
(>95 % dip in U
IEC 61000-4-11
for ½ cycle
40 % U
(60 % dip in U
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 seconds
Magnetic field at the supply
frequency (50/60 Hz) acc. to
3 A/m
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Information on
electromagnetic
Complies
± 4 kV contact discharge
± 8 kV air discharge
Complies
± 2 kV for power supply
lines
± 1 kV for input/output lines
Complies
± 1 kV outer conductor –
outer conductor
± 2 kV outer conductor –
earth
Complies
*
<5 % U
*
T
T
)
(>95 % dip in U
T
for 1/2 cycle
Complies
40 % U
T
T
)
(60 % dip in U
)
T
T
for 5 cycles
Complies
70 % U
T
T
)
(30 % dip in U
)
T
T
for 25 cycles
Complies
<5 % U
T
(>95 % dip in U
for 5 seconds
Complies
3 A/m
Floors should be made of wood, concrete or ceramic
tile. If floors are covered with synthetic material, the
relative humidity must be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
If the user of the device requests continued functioning
T
even when interruptions of the power supply occur, it
is recommended that the device be powered from an
uninterruptible power supply or a battery.
)
T
The magnetic fields at the mains frequency should be at
the values typical of business or hospital environments.