Herunterladen Inhalt Inhalt Diese Seite drucken

Interacoustics Virtual SVV Bedienungsanleitung Seite 46

Inhaltsverzeichnis

Werbung

Virtual SVV™ Instructions for Use - EN
Guidance and manufacturer's declaration - electromagnetic emissions
Virtual SVV™ is intended for use in the electromagnetic environment specified below. The customer
or the user of Virtual SVV™ should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment
and Virtual SVV™.
Virtual SVV™ is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of Virtual SVV™ can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and Virtual SVV™ as recommended below, according to the maximum output power of the communications
equipment.
Rated Maximum output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance
Group 1
Class B
Not Applicable
Not applicable
Separation distance according to frequency of transmitter
[m]
d = 1.17 √ ����
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
Electromagnetic environment - guidance
Virtual SVV™ uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Virtual SVV™ is suitable for use in all commercial,
industrial, business, and residential environments.
d = 1.17 √ ����
80 MHz to 800 MHz
0.12
0.37
1.17
3.70
11.70
Page 42
d = 2.23 √ ����
800 MHz to 2.5 GHz
0.23
0.74
2.33
7.37
23.30

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis