Wand
Weight - 8 ounces/226 grams (with cord)
Length - 9.26 inches/23.5 cm
Breadth -.74 inches/1.9 cm
Power Supply
Output - 9VDC at 2A
Input - 100VAC to 240VAC
Ultradent P/N 5930 VALO GRAND Corded curing light Power
Supply with Universal Plugs
Operating Conditions
Temperature: +10°C to +32°C (+50°F to +90°F)
Relative Humidity: 10% to 95%
Ambient Pressure: 700 hPa to 1060 hPa
Storage and Transport
Device Storage and Transport:
Conditions:
• Temperature: +10°C to +40°C (+50°F to +104°F)
• Relative Humidity: 10% to 95%
• Ambient Pressure: 500 hPa to 1060 hPa
Guidance and Manufacture's Declaration for Electromagnetic Emissions
The VALO GRAND Corded curing light is intended for use in the electromagnetic environment specified below. The customer or user should ensure that it is
used in such an environment.
WARNING: Use only authorized accessories, cables, and power supplies to prevent improper operation, increased electromagnetic emissions or decreased
electromagnetic immunity.
Emissions Test
RF emissions: CISPR 11
RF emission: CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and Manufacture's Declaration for Electromagnetic Immunity
The VALO GRAND Corded curing light is intended for use in the electromagnetic environment specified below. The customer or user should ensure that it is
used in such an environment.
IMMUNITY test
IEC 60601 test level
Electrostatic
± 8 kV contact
discharge (ESD)
± 15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV line to line
IEC 61000-4-5
± 2 kV line to earth
6
Compliance
Electromagnetic environment - guidance
Group 1
VALO GRAND Corded curing light uses a Globtek medical grade 9VDC
adaptor, operates with brown-out protection, and provides limited EMI, RF,
and surge suppression.
Class B
VALO GRAND Corded curing light uses electrical and electromagnetic energy
only for their internal functions. Therefore, any RF emissions are very low
Class A
and are not likely to cause interference in nearby electronic equipment.
The VALO GRAND Corded curing light is suitable for use in all
COMPLIES
establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings for domestic use.
Compliance level
Electromagnetic environment guidance
± 8 kV contact
Physical environment should be restricted to the following:
± 15 kV air
1. IP Code: IP20
2. Do not immerse in liquid.
3. Do not use around flammable gas. Unit is Non-APG and Non-AP.
4. Storage humidity range: 5% - 95%
5. Storage temperatures range: 10° C - 40° C
± 2 kV for power supply lines
Mains power quality should be that of a typical residential, commercial,
Note 1: VALO GRAND
or hospital, or military environment
Corded curing light has no
I/0 ports
± 1 kV line to line
± 2 kV line to earth
Width - .79 inches/ 2 cm
Cord length - 7 feet/2.1 meters
Ratings - CE, IEC 60601
Rating - Medical Grade Class II (UL, CE, RoHS, WEEE)
Cord length - 6 feet/1.8 meters
VALO GRAND Corded curing light power supply provides isolation from
MAINS power