Duty Cycle: The VALO GRAND Corded curing light is designed for short-term operation. At maximum ambient temperature (32°C) 1 minute ON
back-to-back cycling, 30 minutes OFF (cooling-off period).
Accessory Bracket Instructions
1- Bracket should be mounted to a flat, oil-free surface.
2- Clean surface with rubbing alcohol.
3- Peel backing off of the bracket's adhesive tape.
4- Position bracket so the handpiece lifts upward when removed. Press firmly into place.
Maintenance and Cleaning
VALO GRAND Corded curing light is a sealed unit with a hard surface and a scratch resistant glass lens. After each use, moisten a gauze or soft cloth
with an approved surface disinfectant and wipe the surface and lens.
Routinely check the lens for cured dental resins. If necessary, use a non-diamond dental instrument to carefully remove any adhered resin.
The barrier sleeve helps prevent cross contamination and helps keep dental composite material from adhering to the surface of the lens and wand body.
Light meters differ greatly and are designed for specific light guide tips and lenses. Ultradent recommends checking VALO GRAND Corded curing
light in Standard Power mode. NOTE: the true numeric output will be skewed due the inaccuracy of common light meters and the custom LED pack
VALO GRAND Corded curing light uses.
PRODUCT SAFETY WARNINGS:
• DO NOT autoclave.
• DO NOT immerse in any kind of ultrasonic bath, disinfectant, cleaning solutions, or any kind of liquid.
• PRECAUTION: This unit may be susceptible to strong magnetic or static electric fields, which could disrupt the programming. If you suspect this
has occurred, unplug the unit momentarily and then replug it into the outlet.
• DO NOT wipe down the VALO GRAND Corded curing light with caustic or abrasive cleaners. See lists of acceptable cleaners below:
ACCEPTABLE CLEANERS:
• Lysol Brand III Disinfectant Spray (Recommended)
• Isopropyl alcohol
• Ethyl alcohol-based cleaners
• Lysol
®
* Concentrate (alcohol-based only)
• Cavicide™* products (Non-Bleach)**
DO NOT USE:
• Strong alkali detergent of any type, including hand soaps and dish soaps
• Bleach-based cleaners (e.g. - Clorox
• Hydrogen Peroxide based cleaners
• Abrasive Cleansers (e.g. – Comet Cleanser
• Acetone or hydrocarbon based cleaners
• MEK (Methyl Ethyl Ketone)
• Birex
*
®
• Gluteraldehyde
• Quaternary Ammonium Chloride salt-based cleaners (except Cavicide)
• Cavicide1 solution or wipes
* Trademark of a company other than Ultradent.
** If used, it may fade the color
4