Data
Interfaces
Interface/Protocol..............................................................................................................................................RS485/BMS
Connection.....................................................................................................................................................terminals A/B
Max. cable length.............................................................................................................................................. ≤ 1200 m
Recommended cable (screened, connect one side of the screen to PE) ......................................at least J-Y(St)Y 2 x 0,6
Terminating resistor ..................................................................................................................................120 Ω (0.25 W)
Switching elements
Switching elements ......................................................................................................................... changeover contacts
Operating principle, adjustable.......................................................................................... N/C operation / N/O operation
Factory setting ............................................................................................................................................... N/C operation
Number of cycles............................................................................................................................ 12000 operating cycles
Rated contact voltage..........................................................................................................................AC 250 V / DC 300 V
Limited making capacity..................................................................................................................................... AC/DC 5 A
Limited breaking capacity ........................................................................................................ 2 A, AC 230 V, cos phi 0.4
................................................................................................................................................0.2 A, DC 220 V, L/R=0.04 s

General data

EMC immunity..................................................................................................................................acc. to IEC 61000-6-2
EMC emission ..................................................................................................................................acc. to IEC 61000-6-4
Shock resistance acc. to IEC 6068-2-27 (device in operation) ...................................................................... 15 g / 11 ms
Bumping acc. to IEC60068-2-29 (transport) ....................................................................................................40 g / 6 ms
Vibratory resistance acc. to IEC 60068-2-6 (device in operation)........................................................ 1 g / 10 ... 150 Hz
Vibratory resistance acc. to IEC 6068-2-6 (device not in operation).................................................... 2 g / 10 ... 150 Hz
Ambient temperature, operation............................................................................................................. -10 ºC ... +50 ºC
Ambient temperature, storage ................................................................................................................ -40 ºC ... +70 ºC
Climatic category acc. to DIN IEC60721-3-3 ...............................................................................................................3K5
Operating mode ................................................................................................................................................. continuous
Position ............................................................................................................................................................. any position
Connection.................................................................................................................................................. screw terminals
Power circuit:
Cage clamp terminal ............................................................................................................. 1.5 ... 16 mm
Tightening torque, terminal screws................................................................................... 0.5 ... 0.6 Nm (4.3 ... 5.3 lb-in)
Control section:
Cage clamp terminals ................................................................................................................................0.08 ... 2.5 mm
Degree of protection internal components (DIN EN 60529).......................................................................................IP30
Degree of protection terminals (DIN EN 60529) .......................................................................................................... IP20
Installation on standard distributor LTIC107E ...................................................................................1 field / 3 segments
Installation on standard distributor LTIC107D ...................................................................................1 field / 4 segments
Flammability class .................................................................................................................................................UL94V-0
Protection class............................................................................................................................................................... SK I
Power consumption of monitoring module LTIC107E / D............................................................................. < 10 / 14 W
Weight ......................................................................................................................see table "Dimensions and weights"