Herunterladen Inhalt Inhalt Diese Seite drucken

Function Control; General Function - Nouvag Vacuson 40 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Vacuson 40:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Vacuson 40/60

5.8 Function Control

After the Vacuson 40/60 has been set up, check the different operation functions in order to receive
optimum performance and service from the device.

General Function:

1)
Press "ON/OFF" Mains Switch to start the unit
2)
Both Fans on the left and right of the unit will run.
Suction Pump without Pedal:
1)
Press "ON/OFF" Mains Switch to start the unit
2)
Both Fans on the left and right of the unit will run.
3)
Suction Pump is ON.
Keep hand in front of the Air Exhauster „EXHAUST" (Control Unit back panel) → Air Flow
4)
exhausts.
5)
Turn Vacuum Regulator "VACUUM" completely right
6)
Verify Suction Intensity on Cannula's end
7)
Turn Vacuum Regulator "VACUUM" completely left
8)
Verify Suction Intensity on Cannula's end
9)
Switch Off Control Unit with Mains Switch "ON/OFF
Suction Pump with „ON/OFF" – Pedal (optional):
1)
Press "ON/OFF" Mains Switch to start the unit
2)
Both Fans on the left and right of the unit will run.
Keep hand in front of the Air Exhauster „EXHAUST" (Control Unit back panel) → Air Flow
3)
exhausts.
4)
Switch On Suction Pump with „ON/OFF"- Foot Pedal.
5)
Turn Vacuum Regulator "VACUUM" completely right
6)
Verify Suction Intensity on Cannula's end
7)
Turn Vacuum Regulator "VACUUM" completely left
8)
Verify Suction Intensity on Cannula's end
9)
Switch Off Suction Pump with „ON/OFF"- Foot Pedal.
10) Switch Off Control Unit with Mains Switch "ON/OFF
Suction Pump with AIR-Pedal (optional):
1)
Press "ON/OFF" Mains Switch to start the unit
2)
Both Fans on the left and right of the unit will run.
3)
Suction Pump is ON.
Keep hand in front of the Air Exhauster „EXHAUST" (Control Unit back panel) → Air Flow
4)
exhausts.
5)
Turn Vacuum Regulator "VACUUM" completely right
6)
Verify Suction Intensity on Cannula's end
7)
Push Foot Pedal
8)
Release Foot Pedal.
9)
Turn Vacuum Regulator "VACUUM" completely left
10) Verify Suction Intensity on Cannula's end
11) Switch Off Control Unit with Mains Switch "ON/OFF
Troubleshooting:
See capitel 8. Troubleshooting
8
Suction Output will decrease.
Switch "ON/OFF" will illuminate.
Switch "ON/OFF" will illuminate.
Maximum Suction Output.
Pump suction increases
Minimum Suction Output.
Pump suction decreases.
Switch "ON/OFF" not illuminated.
Switch "ON/OFF" will illuminate.
Maximum Suction Output.
Pump suction increases
Minimum Suction Output.
Pump suction decreases.
Switch "ON/OFF" not illuminated.
Switch "ON/OFF" will illuminate
Maximum Suction Output.
Pump suction increases.
Minimum Suction Output.
Pump suction decreases.
Switch "ON/OFF" not illuminated.
NOUVAG AG
No. 31791 04/08

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Vacuson 60

Inhaltsverzeichnis