Herunterladen Inhalt Inhalt Diese Seite drucken

Operation With Two Secretion Jars; Function Control - Nouvag Vacuson 60 LP Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Vacuson 60 LP:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

5.8 Operation with two Secretion Jars

Secretion jars can be hung left and right on the Control Unit. When the first Secretion Jar is full, the
Vacuum Pump must be turned off and the connection tubes attached to the second Secretion Jar
before operation can be resumed.
Even after closing, some liquid will still run into the Jar, subject to the negative
pressure remaining in the system.

5.9 Function Control

After the Vacuson 60LP has been set up, check the different operation functions in order to receive
optimum performance and service from the device.
General Function:
Press "ON/OFF" Mains Switch to start the unit  Switch "ON/OFF" will illuminate.
1)
2)
Infiltration Pump, Suction Pump and Electronicmotor should start only by pressing the Mains
Switch, without operating the Floor Pedal.
3)
Both Fans on the left and right side of the unit will run.
Infiltration Pump:
1)
Turn Pump Regulator ("PUMP") on the Control Unit to the right (adjust infiltration maximum
speed).
2)
Push Foot Pedal down to start Infiltration Pump.
With the Foot Pedal, Infiltration speed can be variably controlled  Pump runs faster when Foot
3)
Pedal is pressed.
4)
Press Foot Pedal completely and vary Infiltration speed with Pump Regulator "PUMP" (on
Control Unit)  Pump runs faster when Pump Regulator is turned right.
5)
Turn Pump Regulator "PUMP" back to right.
 For more information see Chapter 5.2 Floor Pedal Functions and Chapter 5.3 Regulation of the
Infiltration Procedure.
Suction Pump and Suction Interruption
1)
Press Foot Button "VACUUM" (blue) to switch Pump on.
Keep hand in front of the Air Exhauster "EXHAUST" (Control Unit back panel)  Air Flow
2)
exhausts.
Turn Vacuum Regulator "VACUUM" left  Maximum Suction Output.
3)
Verify Suction Intensity on Cannula's end  Pump suction increases.
4)
Press Valve Button on Hand grip  Air will be aspirated by the valve.
5)
Verify Suction Intensity on Cannula's end  nonexistent intake pressure.
6)
7)
Release Valve Button on Hand grip.
Turn Vacuum Regulator "VACUUM" right  Minimum Suction Output.
8)
Verify Suction Intensity on Cannula's end  Pump suction decreases.
9)
10) Foot Button "VACUUM" (blue) switches Suction Pump OFF.
 For more information see Chapter 5.2 Floor Pedal Functions and Chapter 5.4 Regulation of the
Suction Procedure.
Electronicmotor:
Foot Button "MOTOR" (yellow) switches Electronicmotor ON  Vibration movement on Suction
Cannula.
 For more information see Chapter 5.2 Floor Pedal Functions.
Troubleshooting:
See Chapter 8 Troubleshooting.
No. 31793 15/11
Vacuson 60LP
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis