Operation

7.1

Switching the device on and off

The power switch "I/O" (at the rear) is used to switch the control unit on and off.
The device can be switched off at any time irrespective of any procedure for switching off the device.
7.2

Overview: Control elements on the operation panel

A
A) Display: Shows the operating values (See chapter "7.3 Overview: Standard display").
B) "Handpiece" key: Selection of the required handpiece with the corresponding settings.
The factory default settings contain the handpieces 1 : 1, 4 : 1, K-Wire as well as micro saws.
C) "Program" key: Saving individual settings.
4 memory slots are available for individual settings.
By pressing both „HP + PR" keys at the same time the programs can be reset to the factory default settings.
D) "Speed" keys:
Restrict the maximum speed that can be selected using the pedal.
"▲": increases the maximum speed
By pressing both "Speed" keys at the same time "Speed ▲ + ▼" the handpiece testing procedure will be
started. (refer to "7.4.2 Handpiece testing").
E) "Torque" keys:
Restricts the maximum torque.
"▲": increases the maximum torque
With the use of the micro saws and the K-Wire handpiece the torque is fix and cannot be altered.
F) "Pump" keys:
Adjusting the maximal pump flow rate, recalled by the pedal.
"▲": increases the maximum supply quantity
By pressing both "Pump" keys at the same time "Pump ▲ + ▼" the pump will be put on call, pressing again
will switch it off.
OFA-Drill operating instructions, ref. 31595, V49/15
"▼": reduces the maximum speed
"▼": reduces the maximum torque
"▼": reduces the maximum supply quantity