Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data And Specifications - Knauer Smartline 5000 Serie Handbuch

Pumpen
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Vacuum Degasser Unit

Technical Data and Specifications

Table 4
Technical Data and Specifications
Specification:
Dimensions
Weight
Channels
Degassing Flow Path I.D.
Degassing Process
Maximum Flow Rate
Degassing Capacity
Degassing Efficiency
Degassing Channel Pressure
Rating
Dead Volume
Optimized Flow Rate (for
isocratic or gradient formed
50:50 methanol/water)
Materials contacting solvents
Solvent Applicability
Power:
Power Supply
Operating Conditions:
Ambient Temperature
Ambient Relative Humidity (RH) 20 to 80 % RH (without condensation)
Altitude
Indoor vs. Outdoor Use
Pollution Degree
Storage Conditions:
Ambient temperature
Ambient Relative Humidity
Description:
226 x 135 x 390 mm (W x H x D)
4.8 kg (complete Smartline Manager 5000)
1-4 independent / 1-2 independent
1.14 mm analytical version /
1.77 mm semipreparative version
Gas permeation through a fluoropolymer
membrane
10 ml/min analytical / 50 ml/min
semipreparative
~2 ppm at 1 ml/min for analytical version
< 0.5 ppm residual disolved O
70 PSIG
~480 µl per channel for analytical version
~7.7 ml per channel for semipreparative
version
3 ml/min analytical version / 30 ml/min
preparative version
PEEK, Glass-filled PTFE, Teflon AF
universal, except hydrochloric acid
and halogenated carbohydrates
especially hexafluoro isopropanol (HFIP)
via Smartline Pump 1000
90-260 V, 47–63 Hz
10 to 35 ºC
0 to 2000 Meters
Indoor
2
-20 to +60 ºC
20 to 80% RH (without condensation)
13
at 1 ml/min
2
®

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis