This is a Safety Class I product. It must be attached to a grounded power source. To avoid damage to the Analyzer and
the chance of injury, make sure the Analyzer is connected to a power line of the proper voltage.
Any operation requiring access to the inside of the Analyzer, including installation of maintenance items, could result in
injury. To avoid potentially dangerous shock, turn off power and disconnect from the power supply before opening the
Analyzer.
The UV lamp and the display screen contain mercury and may be considered hazardous material in your local area.
Dispose of these items in accordance with federal, state, or local government regulations.
Should the UV lamp become broken or damaged it should be handled in accordance with your organization's toxic waste
handling procedure and disposed of in accordance with federal, state, or local government regulations.
To protect against accidental exposure to ultra-violet radiation, do not operate the UV lamp outside of its protective
housing. The UV lamp housing may be hot; use extreme caution whenever touching the UV lamp housing.
Always stop analysis before turning off the Analyzer.
This is a Class A product. In a domestic environment, this product may cause electromagnetic interference in which case
the user may be required to take adequate measures to correct the interference.
To avoid false TOC readings and possible damage to the Analyzer, always make sure the sample inlet is open and the DI
water cartridge is filled before starting analysis. Always close the door of the Standard iOS System when not sampling
from a vial.
Make sure the DI water cartridge is full, particularly when running samples with high TOC or high salt concentrations.
Always "clean-up" the Analyzer by running low-TOC DI water after running high TOC or salt samples.