40 MHz
10 kHz
7.9 Temperature ranges °C
With temperature sensor type K and sensor adapter.
Measuring range
- 20 °C to approx. 0 °C
0 °C to approx. 100 °C
101 °C to approx. 500 °C
501 °C to approx. 800 °C
7.10 Temperature ranges °F
With temperature sensor type K and sensor adapter.
Measuring range
- 4 °F to approx. 32 °F
33 °F to approx. 212 °F
213 °F to approx. 932 °F
933 °F to approx. 1472 °F
7.11 PEAK HOLD
DC/ AC V Measuring range
1000 V/ 750 V
DC/ AC A Measuring range
8.

Measuring with the BENNING MM 7

8.1 Preparation for measurement

Store and use the BENNING MM 7 only under the correct temperature condi-
tions stated. Always avoid longer exposure to sunlight.
-
Check the rated voltage and rated current stated on the safety test leads.
The safety test leads supplied with the unit are suitable for the rated voltage
and current of the BENNING MM 7.
-
Check the insulation of the safety test leads. If the insulation is damaged in
any way, do not use the leads.
-
Check the continuity of the safety test leads. If the conductor in the safety
test lead is interrupted, do not use the leads.
-
Before selecting another function with the rotating switch , always discon-
nect the safety test leads from the measuring point.
-
Sources of strong current in the vicinity of the BENNING MM 7 may cause
unstable or incorrect readings.
8.2 Voltage and current measurement
Always observe the maximum voltage to earth potential!
Electrical hazard!
The maximum voltage which may be applied to the sockets
-
COM- socket 
-
socket for V, Ω, Hz, ºC, ºF,
-
socket for mA range  and the
-
socket for 10 A range 
of the BENNING MM 7 to earth is 1000 V.
11/ 2008
± (0.01 % of reading + 1 digit)
Measurement accuracy
± (2 % of reading + 4 °C)
± (1 % of reading + 3 °C)
± (2 % of reading + 3 °C)
± (3 % of reading + 2 °C)
Measurement accuracy
± (2 % of reading + 8 °F)
± (1 % of reading + 6 °F)
± (2 % of reading + 6 °F)
± (3 % of reading + 4 °F)
400 mV
4 V
40 V
400 V
40 mA
400 mA
10 A
BENNING MM 7
Measurement accuracy
± (1.5 % of reading + 300 digits)
± (1.5 % of reading + 60 digits)
± (1.5 % of reading + 60 digits)
± (1.5 % of reading + 60 digits)
Measurement accuracy
± (3.0 % of reading + 60 digits)
± (3.0 % of reading + 60 digits)
± (1.5 % of reading + 60 digits)
200 kHz
Overload protection
600 V
eff
600 V
eff
600 V
eff
600 V
eff
Overload protection
600 V
eff
600 V
eff
600 V
eff
600 V
eff
not specified
1 V
eff
16