Continuity Test; Measuring With The Benning Cm - Benning CM 1-4 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7.3 AC current ranges
Measuring
range
200 A
7.4 Resistance measuring ranges
Measuring
range
9999 Ω
50.00 kΩ
Measuring voltage: approx. 0.5 V

7.5 Continuity test

The integrated buzzer sounds at a resistance R lower than approx. 1.8 kΩ to
2.7 kΩ.
Acoustic indication: signal of 2.7 kHz
Optical indication:
Response time:
Open-circuit voltage: approx. 0.5 V
Overload protection: 1000 V
7.6 Single-pole external conductor test (phase indication)
Sensitivity:
Frequency range:
Acoustic indication: signal of 2.7 kHz
Optical indication:
7.7 Phase sequence test
Sensitivity:
Frequency range:
Optical indication:
8.

Measuring with the BENNING CM 1-4

8.1 Preparation for measuring
Operate and store the BENNING CM 1-4 at the specified storage and operating
temperatures only! Do not permanently expose the device to sunlight.
-
Check stated nominal voltage and nominal current on the safety measuring
leads. Nominal voltage and current of the enclosed safety measuring leads
comply with the respective values of the BENNING CM 1-4.
-
Check insulation of the safety measuring leads. If the insulation is damaged,
the safety measuring leads must be replaced immediately.
-
Check the safety measuring leads for continuity. If the conductor in the safety
measuring lead is interrupted, replace the safety measuring leads immedi-
ately.
-
Attach the protective caps (CAT III/IV) to the contact tips before making
measurements in circuits of overvoltage category CAT III or IV.
see fig. 1b: Using the protective caps
-
The rear holder of the BENNING CM 1-4 is intended for attaching the mea-
suring leads for voltage measurement and when not in use/for storage.
see fig. 1c: Measuring leads holder
-
Strong sources of interference in the vicinity of the BENNING CM 1-4 might
involve unstable readings and measuring errors.
8.2 Self-testingfunction(self-test)
After each switch-on via the Ω/A key 6 or the HOLD key 7 , the BENNING CM 1-4
carries out a self-test. An acoustic signal is emitted and all display segments and
LED indications must light up briefly.
If the self-test detects an abnormal behaviour, the device must be switched off
immediately and secured against unintended operation.
Before and after using the BENNING CM 1-4, check the function of the voltage
measurement and the single-pole external conductor test (phase indication) at a
known voltage source.
09/ 2019
Resolution
in frequency range 45 Hz - 65 Hz
0.1 A
Resolution
1 Ω
0.01 kΩ
-LED
<100 ms
AC/DC
90 V to 1000 V (voltage to earth) in earthed mains
45 Hz - 65 Hz
-LED
90 V to 1000 V (voltage to earth) in earthed mains
45 Hz - 65 Hz
-LED,
BENNING CM 1–4
Measurement accuracy
± (3.0 % of reading + 7 digits)
Measurement accuracy
± (1.5 % of reading + 7 digits)
-LED
Overload
protection
200 A
Overload
protection
1000 V AC/DC
25