8. Transferring measurements

Connect the blood pressure monitor to your PC using the USB
cable.
No data transfer may be launched whilst performing a
measurement.
PC is shown on the display. Begin the data transfer
in the "beurer HealthManager" PC software. During
the data transfer, an animation is shown on the
display. A successful data transfer is displayed as
in figure 1. If the data transfer is unsuccessful, an
error message appears as in figure 2. In this case,
interrupt the PC connection and start the data
transfer again.
After 30 seconds of not being in use or if commu-
nication with the PC is interrupted, the blood pres-
sure monitor switches itself off automatically.
9. Cleaning and storing the device and
cuff
• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.

10. Error messages/trouble-shooting

In case of faults, the
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
appears on the display)
• systolic or diastolic pressure was outside the measurement
range ( or Lo appears on the display)
• the cuff is fastened too tightly or loosely (
on the display)
• the pump pressure is higher than 300 mmHg (
on the display)
figure 1
• pumping up takes longer than 160 seconds (
the display)
• there is a system or device error (
pears on the display)
• the batteries are almost empty
• the data could not be sent to the PC (
figure 2
display).
In such cases, repeat the measurement. Ensure you do not
move or speak. If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical specifica-
tions", the technical alarm will appear on the display indicating
either "     " or " Lo  " . In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
30