Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The BM 55 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 55 should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
± 6 kV contact
(ESD) IEC 61000-4-2
± 8 kV air
± 2 kV for power supply
Electrical fast
lines
transient/burst
± 1 kV for input/output
IEC 61000-4-4
lines
± 1 kV line(s) to line(s)
Surge IEC 61000-4-5
± 2 kV line(s) to earth
<5 % U
for 0.5 cycle
Voltage dips, short inter-
40 % U
ruptions and voltage
for 5 cycles
variations on power sup-
70 % U
ply input lines
for 25 cycles
IEC 61000-4-11
<5 % U
for 5 s
Power frequency
(50/60 Hz) magnetic field
3 A/m
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
(>95 % dip in U
)
<5 % U
T
T
for 0.5 cycle
(60 % dip in U
)
40 % U
T
T
for 5 cycles
(30 % dip in U
)
70 % U
T
T
T
for 25 cycles
(>95 % dip in U
)
<5 % U
T
T
for 5 s
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
(>95 % dip in U
)
Mains power quality should be that of a typical com-
T
T
mercial or hospital environment. If the user of the BM 55
(60 % dip in U
)
requires continued operation during power mains inter-
T
T
ruptions, it is recommended that the BM 55 be powered
(30 % dip in U
)
from an uninterruptible power supply or a battery.
T
(>95 % dip in U
)
T
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.