Herunterladen Diese Seite drucken

HBM WTX110 Bedienungsanleitung Seite 177

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

The linearization software supports up to 6 correction points, though in practice
1 or 2 are almost always enough. The linearization points should be placed at
the points on the weight curve where the greatest variations occur.
Linearization points must be lower than the set maximum load. The lines
between adjacent linearization points are calculated as straight, meaning the
variations are interpolated in linear mode.
Menu
Select group 1-9
→ Linearization points already present
Fixpoint 1:
Calibrate Fixpt?
Lin.Signal
→ Clr key pressed in Fixpoint n step
Delete Fixpoint n? N
Information
After deleting a linearization point, the remaining points are renumbered in
ascending order.
WTX110
Key
999999
N
0.50000
A4774-1.0 HBM: public
Function/Secondary selection
Press 3 x for group 3 Linearization
Display existing linearization points (1 to max. 6)
Delete existing linearization point
Linearization signal is displayed
Delete linearization point; continue to New
Y
Fixpoint? step
Do not delete linearization point; continue
N
to next linearization point
Calibration
175

Werbung

loading