9.3
Emergency release
Fig. 22: Emergency release
1
Hole
10
Disposal
10.1
General instructions
AB1401en
2.
Wait 10 seconds.
3.
Set the mains switch to
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
Risk of electric shock due to maintenance and servicing
work on live device.
WARNING
Danger of cutting and crushing due to moving rotor.
Personnel:
Trained user
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
2.
Insert the hex key horizontally into the hole (
until the lid opens.
3.
Remove the hex key from the hole (
1
4.
Check whether the left side of the
power is restored.
When the left side of the
[STOP/OPEN] button so that the motorised lid lock assumes the
home position (open) again.
The device can be disposed of via the manufacturer.
A Return Material Authorisation (RMA) form must always be
requested for a return.
If necessary, contact the Technical Service Department of
the manufacturer:
Rev.: 13 / 11.2023
[I] .
WARNING
Disconnect the device from the mains before carrying
out repairs and maintenance.
Do not open the lid until the rotor has stopped.
1  ).
[STOP/OPEN] button flashes when
[STOP/OPEN] button flashes, press the
Andreas Hettich GmbH & Co. KG
Föhrenstraße 12
78532 Tuttlingen, Germany
Phone: +49 7461 705 1400
E-Mail: service@hettichlab.com
Disposal
1  ) and turn clockwise