Operation
6.7.2
Centrifugation with time preselection
6.7.3
Short-time centrifugation
6.8
Quick stop function
30 / 44
Personnel:
■
Trained user
1.
Set centrifugation parameters or retrieve a program.
[START/IMPULS] button.
2.
Press the
The centrifugation run is started.
The
[START] button lights up during the centrifugation run.
The rotor speed or the RCF value, the temperature in the cen-
trifuging chamber (only for centrifuges with cooling) and the
remaining time are displayed during the centrifugation run.
3.
Ramp-down takes place with the selected brake level after the time
has elapsed or if the centrifugation run is cancelled.
The brake level is displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
The right side of the
centrifuge is in ramp-down.
The left side of the
at a standstill.
The light on the
[STOP/OPEN] button go out.
Personnel:
■
Trained user
1.
Press and hold the
[START/IMPULS] button lights up during the centrifugation
The
run.
Timing starts at 00:00.
The rotor speed or the RCF value, the temperature in the centri-
fuging chamber (only for centrifuges with cooling) and the elapsed
time are displayed during the centrifugation run.
[START/IMPULSE] button to end the centrifugation run.
2.
Release the
Ramp-down takes place with the set brake level. The brake level is
displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
Personnel:
■
Trained user
[STOP/OPEN] button twice.
Press the
Ramp-down with brake level "9" (shortest ramp-down time) is dis-
played and executed.
Rev.: 13 / 11.2023
[STOP/OPEN] button lights up when the
[STOP/OPEN] button lights up when the rotor is
[START/IMPULS] button and the right side of the
[START/IMPULS] button.
AB1401en