Adapter with positioning pin
1
Fig. 18: Adapter with positioning
pin
1
Inserting
2
Positioning pin
3
Inserting
the groove
removing
6.5
Loading
Filling centrifuge tubes
AB1401en
2
3
Insert the adapter (
The positioning pin (
Remove the adapter (
Personnel:
Trained user
Rev.: 13 / 11.2023
1  ) into the bucket
2  ) must be in the groove ( 3  ) of the bucket.
1  ) vertically upwards out of the bucket.
WARNING
Risk of injury from contaminated sample material.
Contaminated sample material escapes from the sample
tube during centrifugation.
Use centrifuge tubes with special screw caps for haz-
ardous substances.
For risk group 3 and 4 materials, use a biosafety system
in addition to the sealable centrifuge tubes (see WHO's
'Laboratory Biosafety Manual').
NOTICE
Damage to the device due to highly corrosive substances.
Highly corrosive substances may impair the mechanical
strength of rotors, buckets and accessories.
Do not centrifuge highly corrosive substances.
Standard glass centrifuge tubes can be loaded up to RCF
4000 (DIN 58970 part 2).
Operation
25 / 44