Concerned parts Processing
Temperature: 20°C~40°C
or
Automated cleaning: Carefully put the components into the
washer-disinfector on a tray and set the parameters as follows,
then start the program:
• 4 min pre-washing with cold water (<40°C);
• emptying
• 5 min washing with a mild alkaline cleaner at 55°C;
• emptying
• 3 min neutralizing with warm water (>40°C);
• emptying
• 5 min intermediate rinsing with warm water (>40°C);
• emptying
The automated cleaning processes have been validated by using
0.5% neodisher MediClean forte (Dr. Weigert).
Concerned parts Processing
Contra-angle
Automated Thermal Disinfection in washer/disinfector under
PN:51200158
consideration of national requirements in regards to A0 value
(see EN ISO 15883).
Silicon sleeve
A disinfection cycle of 5 min disinfection at 93°C has been
PN:51200161
validated for the device to achieve an A0 value of 3000.
After manual cleaning, the instrument should be automated
disinfected or sterilized immediately. A manual disinfection is not
recommended.
Concerned parts Processing
Contra-angle
Automated Drying:
PN:51200158
Drying of outside of instrument through drying cycle of washer/
disinfector. If needed, additional manual drying can be performed
Silicon sleeve
through lint free towel. Insuffl ate cavities of instruments by using
PN:51200161
sterile compressed air.
Concerned parts Processing
Contra-angle
Visual inspection for cleanliness of the components and
PN:51200158
reassembling. Functional testing according to instructions of
use. If necessary, perform reprocessing process again until the
Silicon sleeve
component is visibly clean.
PN:51200161
Before packaging and autoclaving, make sure that the device has
been maintained acc. to the manufacturer's instruction.
Lubrication: Attaching the spray nozzle to oil can and contra
angle, press the oil can button more than 3 seconds, till all the
black oil fl ow out from the head of the contra angle.
2 - Disinfection
3 - Drying
4 - Functional Testing, Maintenance:
Warning
• Make sure the contra angle is dry before
moving to the next step.
Warning
Warning
Warning
• Before autoclaving, the contra angle must
be lubricated.
• Only the contra angle needs to be
lubricated.