Herunterladen Diese Seite drucken

Coltene MicroMega Dual Move Bedienungsanleitung Seite 6

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
• Please set the parameters according to the fi le manufacturer's recommendations.
• The use of torque reverse function can effectively protect the fi le from separation within the canal.
• If the auto reverse is too frequent during use, please recapitulate, irrigate and lubricate the root
canal or increase the torque as per fi le manufacturer's recommendations.
9. 4. Additional settings (Fig. 9.G)
1 - When the device is powered off, press simultaneously the S and the main buttons.
2 - Touch S to select the advanced parameter to be set.
3 - Touch < / > (- / +) to set values.
4 - Press the main switch to escape this mode and come back to the standby screen.
Volume of beep (BEEP VOLUME)
9.G.1
Low (Low), middle (Mid) or high (High) volume of
audible signals. Factory setting: Middle (Mid).
Auto Power off (AUTO P.W.R)
9.G.2
Automatic power off after 3, 5, 10 or 15 min.
Factory setting: 10 min.
Orientation of the display (Hand)
9.G.3
Right (RIGHT) or left (LEFT) orientation.
Factory setting: RIGHT Hand.
When connected to Dual Pex apex locator, Dual Move offers advanced functions, see section 10 to 13.
9. 5. Calibration (Fig. 9.H)
Connect the contra angle to the Dual Move handpiece. Do not insert a fi le.
Enter the calibration interface (see section 9.4) then choose 'ON'.
9.H.1
Press the main switch to start the calibration.
9. 6. Error codes
The battery is fl at, auto power off will occur. Charge
9.I.1
the device immediately.
9.I.2
Motor overload. Reduce the load.
• Follow the instructions of the fi le manufacturer to set the functions and parameters.
• Calibration is strongly recommended after 10-20 autoclaving cycles for the contra-angle and after
its replacement.
• Calibration shall be performed with a 50% battery level as a minimum.
• Do not apply pressure to the contra-angle during calibration.
• If an error code is displayed, please contact your after-sales service.
Refer to the Dual Pex user manual to install it correctly before connecting to Dual Move.
10. Dual Pex to Dual Move connection (Fig. 10)
1 - Make sure that Dual Move is in standby mode.
6
connected to
Inactivity time (SET TIME)
9.G.4
Automatic return to the standby display after the
specifi ed time of inactivity: 1-15s. Factory setting: 10s.
Calibration (Calibration)
9.G.5
Automatic calibration of speed and torque. See next
section.
Bars show the progress.
9.H.2
2 beeps when completed.
Long use time, risk of overheating. Stop using the device
9.I.3
for a while.
(apex locator)

Werbung

loading