Installation
20
Instructions for Use Sartocon
5.3

Installing the Cassette Into the Filter Holder

WARNING
Danger of personal injury and property damage caused by incorrect
installation of the cassettes!
Ensure the correct orientation of the connectors and slots when installing
the cassette (see chapter "4 Components Filter Device", page 9).
WARNING
If you use filter holders (clamping elements) from other manufacturers,
different clamping forces may be required!
Ensuring safety when using filter holders from other manufacturers is the
sole responsibility of the user of these filters. Improper handling with filter
holders from other manufacturers will void the filter warranty.
IMPORTANT
Observe the instructions for the Sartocon
trapezoidal threads, PEEK washers and bronze clamping nuts:
— Replace the clamping nuts and the trapezoidal threads if they show
wear. Please refer to the user manual for the filter holder.
— Before each use, the PEEK plain washers must be cleaned with water in
order to completely maintain the functionality of the filter holder. After
every 10 uses, the plain washers should be replaced with new ones.
— Keep the thread of the pull rods and tensioning nuts clean and, above
all, free of grease! If the threading becomes contaminated with grease
or dirt, the tensioning forces will be affected and the cassettes may be
destroyed.
Proceed as follows to clamp the cassette in the filter holder:
Remove the cassette from the packaging.
t
Check again that the cassette type is correct.
t
— If the Sartocon
Slice or Sartocube
®
installed into a different holder than currently supplied by Sartorius
Stedim Biotech, please make sure to check required compression of the
holder with a clamp force measuring device. To avoid cassette damage,
please contact our application specialist.
— Preferably a Sartorius TFF System should be used with Sartocon
Sartocube
Self Contained cassettes. If a system with a manual holder is
®
used, please check the clamp force on a regular basis.
Slice Self Contained | Sartocube
®
Slice filter holder with
®
Self Contained TFF cassette is
®
Self Contained
®