Integrity Test

Each cassette is fully tested at Sartorius. It is therefore recommended that
no other integrity test be carried out on-site.
Cassette configuration
Hydrosart
Sartocon
®
®
with 0.1 to 0.42 m² filtration area
PES Sartocon
Slice Self Contained
®
with 0.1 to 0.42 m² filtration area
Hydrosart
and PES Sartocube
®
with 0.7 to 3.5 m² filtration area
Hydrosart
and PES Sartocube
®
with 7 m² filtration area
Hydrosart
and PES Sartocube
®
with 14 m² filtration area
5.4

Information about Complaints

If you would like to send cassettes to Sartorius Stedim Biotech for review,
please ensure the following:
— that the customer service is contacted, who will provide you with a return
authorization form.
— that the cassettes have been completely cleaned;
— that the cassettes have been disinfected;
— that the following information is provided:
— the medium that was filtered using the cassettes;
— any errors that were observed;
— the operating parameters in use at the time of filtration (pressure
settings, operating time, process cycles, cleaning conditions,
operating temperature);
— cleaning method used;
— disinfecting agent used.
Slice Self Contained | Sartocube
®
Slice Self Contained
Self Contained
®
Self Contained
®
Self Contained
®
Self Contained
®
Integrity test limit at
1 bar test pressure
≤ 5 mL/min
≤ 15 mL/min
≤ 50 mL/min
≤ 75 mL/min
≤ 100 mL/min