SP25M user's guide
www.renishaw.com
Test setup criteria
CMM specification
CMM controller
Probe calibration
Data filter used
Artefacts used
Scanning speed
Touch speed
Back off speed
Bi-directional scans
Issued 06 2022
Test site
Styli used
Renishaw UK
Renishaw's M3 stylus range
3
U
= 0.48 µm + L / 1000
Renishaw's UCC2
Renishaw's third order polynomial calibration method
Harmonic simple cut off order = 60 UPR (undulations per revolution)
Ø25 mm ﴾nom﴿ calibration sphere ﴾for the ISO 10360 Part 2 / Part 4
tests)
Ø50 mm ﴾nom﴿ ring gauge ﴾for the ring gauge scan tests﴿
5 mm/s (for the ISO 10360 Part 4 and ring gauge scan tests)
5 mm/s (for the ISO 10360 Part 2 tests)
1 mm/s (for the ISO 10360 Part 2 tests)
The effect of probe misalignment and contact friction will result in
radius changes with scanning direction if a suitable calibration
procedure is not used.
Mechanical hysteresis within a probe mechanism is best
demonstrated by a bi-directional scan.
Therefore, Renishaw have decided to take the high integrity
approach of quoting ring gauge metrology using bi-directional scan
data to demonstrate the performance of the Renishaw probe and
calibration method.
68