SP25M user's guide
www.renishaw.com
8. Measure four points on the top surface of the PA25-20 to create a plane [PLN1]. Translate this plane to the Z-axis and offset this plane by [Z
- length of stylus - 21.25 mm] and set the Z-axis origin to this plane.
Step 8 - take four points on top of PA25-20
Step 9 - take two points on front side faces of PA25-20
Step 10 - take two points on port inside rails of PA25-20
9. Measure two points [P1 and P2] on the front faces of the PA25-20. Construct a line [L1] between these points, rotate this line to the X-axis,
offset the line by [Y + 8.75 mm] and origin the Y-axis position of the line.
10. Measure two points [P3 and P4] on the inside rails of each side of the PA25-20. Construct a point [P5] between them.
11. Create a datum coordinate system [dat_TP20_port(n)] for the TP20 module docking position of the port. This will be at the position of [P5],
the Y position of [L1] and the Z position of [PLN1].
12. Repeat steps 1 to 13 above for all other module combinations required.