SP25M user's guide
www.renishaw.com
Put down routines
Put down routine - SM25 scanning modules and TM25-20 adaptor module
Move description
Clearance position
Enter port
Detach module
Clear of module
Exit port
NOTE: To ensure optimum metrology performance it is recommended that following any module change the probe head is
unlocked and then relocked.
Put down routine - SH25 stylus holders
Move description
Clearance position
Enter port
Detach stylus holder(s)
Clear of module
Exit port
Put down routine - TP20 module
Move description
Clearance position
Enter port
Detach TP20 module
Clear of module(s)
Exit port
* Coordinates assume aligned with Y-axis as per
** Calculate nominal 'Z' docking position according to the stylus length used (see calculation formula in section
position for SH25 stylus
holders).
Issued 06 2022
X-axis (mm)*
[dat_MOD_port(n)]
no change
no change
no change
no change
X-axis (mm)*
[dat_SH_port(n)]
no change
no change
no change
no change
X-axis (mm)*
[dat_TP20_port(n)]
no change
no change
no change
no change
FCR25 mounting instructions
Y-axis (mm)*
[dat_MOD_port(n)] - 30 mm
[dat_MOD_port(n)]
no change
no change
[dat_MOD_port(n)] - 30 mm
Y-axis (mm)*
[dat_SH_port(n)] - 30 mm
[dat_SH_port(n)]
no change
no change
[dat_SH_port(n)] - 30 mm
Y-axis (mm)*
[dat_TP20_port(n)] - 30 mm
[dat_TP20_port(n)]
no change
no change
[dat_TP20_port(n)] - 30 mm
Z-axis (mm)*
[dat_MOD_port(n)]
no change
[dat_MOD_port(n)] + 3 mm
[dat_MOD_port(n)] + 8 mm
no change
Z-axis (mm)*
[dat_SH_port(n)]
no change
[dat_SH_port(n)] + 3 mm
[dat_SH_port(n)] + 8 mm
no change
Z-axis (mm)*
** [dat_TP20_port(n)]
no change
**[dat_MOD_port(n)] + 3 mm
** [dat_MOD_port(n)] + 8 mm
no change
Establishing the docking
45