7. Slowly lower the probe body towards the probe module ensuring that the XY position of the probe body does not collide with the port.
Step 6 and 7 - moving to approximate XY position then refining to ensure good alignment between body and module
8. Continue to lower the probe body towards the probe module until the magnetic attraction between the kinematic joint makes the module
move towards the probe body. The module should move slightly towards the kinematic joint of the probe body without tilting (indicating
good XY alignment). If any tilting occurs when the module moves towards the probe body (indicating poor XY alignment) then the probe head
and body should be repositioned and the process should be repeated.
Step 8 - slowly move body downward in Z axis until module jumps straight upwards due to magnetic attraction
9. Slowly lower the probe body on to the module, stop any movement when the LED on the probe head illuminates. This indicates that an
electrical connection is made.
Step 9 - slowly move body downward in Z axis - STOP immediately when the probe head LED is lit