Initial operation

 A type B residual current circuit breaker must be used if the device is additionally protected with a residual current
circuit breaker in the building installation. When using a different type, the residual current circuit breaker may
either not switch off the unit if there is a fault on the unit, or it may switch off the unit even though there is no fault
on the unit.
 Types 5005-90 and 5005-08 are permanently connected devices. With permanently connected devices, a switch
must be fitted in the building installation to disconnect the mains supply to the device, in accordance with the
laboratory equipment standard EN / IEC 61010-1. The switch must be located near the device, be readily
accessible to the user and be marked as the means of isolation for this device. It must be possible to secure the
switch to prevent it being switched on again.
 Types 5005-80, 5005-90 and 5005-08, may only be installed by authorized skilled personnel.
These devices must be installed according to the enclosed set-up and installation instructions,
AH5005-02EN.
 Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
 Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
 Centrifuge with PE connector:
If required, connect the PE connector at the back of the device with an additional medical potential equalization
system.
 Check whether the mains voltage tallies with the statement on the type plate.
 For types 5005, 5005-50 and 5005-80:
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specification".
 Switch on the mains switch in the building installation.
 Switch on the switch on the front panel. Switch position "".
The following display follows:
1. Type of centrifuge, 2. The last rotor code detected by the rotor identification and the maximum speed of the
rotor (n-max-Rotor), 3.The programme version, 4.
 Open the lid.
The centrifuge data of the last used program or program 1 will be displayed.
AB5005DEENFRIT
OPEN
OEFFNEN.
Rev.18 / 04.2023
EN
55/165