Transport and storage

9.1

Transport

When the device and accessories are transported, the following ambient conditions must be complied with:
Ambient temperature: -20°C to +60°C
Relative humidity: 10% to 80%, non-condensing
9.2

Storage

The device and the accessories may only be stored in closed and dry rooms.
When the device and accessories are stored, the following ambient conditions must be complied with:
Ambient temperature: -20°C to +60°C
Relative humidity: 10% to 80%, non-condensing
10

Initial operation

Only for types 4950-08, 4950-58, 4950-78 and 4950-88:
According to the laboratory device standard EN / IEC 61010-1, there must be a switch installed in the building
installation to disconnect the power to the device.
This switch has to be placed close to the centrifuge, must be easily reachable for the user and has to be
characterized as a separator for this unit. The switch must also be able to be secured against being switched on
again.
Types 4950-70, 4950-78, 4950-80 and 4950-88 may only be installed by authorized skilled personnel.
These devices must be installed according to the enclosed set-up and installation instructions,
AH4950EN.
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Centrifuge with PE connector:
If required, connect the PE connector at the back of the device with an additional medical potential equalization
system.
Centrifuge with optical interface:
Connect the optical centrifuge interface with a light-conducting cable (not included in the scope of delivery) to the
PC.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specification".
Turn on the mains switch. Switch position "Ι".
The following display follows:
1. Type of centrifuge, 2. The last rotor code detected by the rotor identification and the maximum speed of the
rotor (n-max-Rotor), 3.The programme version, 4.
Open the lid.
The centrifuge data of the last used program or program 1 will be displayed.
AB4950DEENFRIT
OPEN
OEFFNEN.
Rev. 10 / 03.2022
EN
51/153