Connector J11 - Purge Flow Meter Connector
Figure 15
Tab. 5
PIN
1-3
5-6
Use this connector (J11) only with the Agilent solid state flowmeter model 969-
CAUTION!
9114 or 969-9115. Any other connection could lead to serious injury.
Turbo-V 700 ICE Controller User Manual / 87-900-904-01 (L)
J11 connectors
DESCRIPTION
Flow meter power supply: 12 Vdc,
10 mA, (pin 1 negative - pin 3 positive)
Analog signal with a voltage value proportional to the flow.
Between 0 and 3 V it is linearly proportional to between 0 and 100 sccm
(pin 6 negative, pin 5 positive)