Technical Information
Use

Pump Shutdown

Press the front panel STOP push-button or remove the remote signal;
the power from the turbopump will be removed and the pump will
begin to slow down.

Power Failure

In the event of a power failure (momentary or long term), the Turbo-
V ICE controller will stop the turbopump and all the interconnected
pumps/devices. The Turbo-V ICE vent valve device, if used, will vent
the turbopump only if the power failure is longer than the preset
delay time. When power is restored, the Turbo-V ICE controller
automatically restarts the interconnected devices and the turbopump
in the proper sequence.
Figure 34
252/262
Cycle diagram
Turbo-V 700 ICE Controller User Manual / 87-900-904-01 (L)