17
Technical Information
Interconnections
Forepump relay coil is an independent user supplied item.
NOTE

Connection J10 - Controller to Pump Connection

A five-meter long cable is provided to connect the controller to the
pump. The following figures show the controller output connector
configuration where pins:
Figure 14
218/262
A-F = upper bearing sensor
B-C-D = 54 Vac 3-phase output to pump motor stator
E = ground
A-G =lower bearing sensor
Controller-to-pump connector
(applicable to model 969-9446 and 969-9546)
Turbo-V 700 ICE Controller User Manual / 87-900-904-01 (L)