Herunterladen Inhalt Inhalt Diese Seite drucken

Medel ELITE Gebrauchsanweisung Seite 79

Vorschau ausblenden Andere Handbücher für ELITE:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Table 2
Immunity Test
Electrostatic discharge (ESD) IEC
61000-4-2
Electrical fasttransient/burst
IEC 61000-4-4
Surge IEC61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4-11
Power frequency magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – electromagnetic Immunity
IEC 60601-1-2
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition
frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
% UT; 0,5 cycle. At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°.0 % UT;
1 cycle and 70 % UT; 25/30 cycles; Single
phase: at 0°.0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between
0,15 MHz and 80 MHz 80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance level
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°, 180°, 225°, 270° and
315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80
MHz 80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
79