InPro 6050 Series O
4.2

Connection

To connect the sensor to the transmitter, a VP cable
should be used. This ensures a secure link between
the transmitter and the sensor under harsh industrial
conditions. The robust watertight IP 68 connector
housing guarantees maximum process safety. The
slit of the connector should be aligned with the pin of
the plug, and the two parts should be then screwed
tightly together.
h
Note: For instructions concerning the connection of
the VP cable to the transmitter, please refer to the
transmitter instruction manual.
4.3

Initial start-up

h
Note: The protective cap at the tip of the sensor
should be removed before putting the sensor into
operation.
When the sensor is installed for the first time or has
been disconnected from the voltage source (polariz-
ation voltage from the transmitter or polarization mo-
dule) for longer than 5 minutes, the sensor must be
polarized before calibration or the first measurements.
This can be achieved by connecting the sensor to
the transmitter or a polarization module for at least
six hours. The recommended polarization voltage is
–675mV.
© 03 / 2023 METTLER TOLEDO
Printed in Switzerland
Sensor
2
VP cable
Plug
Pin
Slit
Connector
InPro 6050
InPro 6050
52 200 251 D
9