6.0 Notes on cleaning and care

6.1

Basic information

Remove the mains plug from the socket before starting
cleaning and disinfection!
The measures described for cleaning and disinfection or
sterilisation are not intended to replace the regulations
applicable for operation in each case!
Always comply on principle with the relevant manu-
facturer's information regarding concentration and notes
for use!
Attention: The lid parts and silicone hoses might get
dyed by some disinfectants; a fact which does not take
effect on the attributes of the materials. These silicone
and collection jar parts may also be boiled (for more than
10 min).
Do not use any
Disinfectants containing organic or inorganic
acids or bases since these may cause corrosion
damage.
Disinfectants containing chloramines, phenol deri-
vatives or anionic tensides since these may cause
stress cracks in the plastics used.
Basically, all parts which come into contact with secretion
must be cleaned, disinfected or sterilized after every
use.
Silicone hoses and parts of the collection jar (jar and lid)
are autoclave safe (up to 136°C). These parts can also be
immersed in commercial disinfectants (see page 12).
We recommend changing the hoses every 4 weeks!