Fig. 2. Rear view of the ATMOS LC 16
Power attachment
Power switch
Fuse support
8
4.0

Setting up and starting up

The ATMOS LC 16 is delivered ready for use.
Lift the unit out of the cardboard.
Check whether the voltage values on the data plate
correspond with the inbuilding voltage.
Prior to first operation, pay attention to the safety infor-
mation in chapter 2.0.
The unit must be allowed to stand for up to six hours at
room temperature prior to starting up for the first time
following transport at temperatures below freezing. The
unit may not be operated if it has not acclimatised as
this might damage its diaphragms.
Join suction hose (
(
, fig. 1, page 4).
Connect silicone hose (
ment (
, fig. 1, page 4) and attachment on the unit (
fig. 1, page 4) (pay attention to green signal ring).
Connect the power cable (
ment and the power plug to the local power socket.
Start the unit with the switch (
Changing fuse (fig. 2)
Loosen the fuse support by compressing the lower spring
clip and pull it out (use a 2 mm flat screwdriver, if necessary).
Replace the fuse (see fuse values in chapter 10.0 Technical
specifications) and insert the fuse support again.