Herunterladen Inhalt Inhalt Diese Seite drucken

GF UD2100 Betriebsanleitung Seite 22

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

UD2100 Ultrasonic Doppler Flow Meter
-- Communication --------
Protocol
Address
BPS
Parity
Stop Bits
22
Communication (Optional)
Modbus
Press
or
001
9600
cursor before each menu item and
Even
store and
again to return to the Main Menu.
1
MODBUS Protocol Information:
Transceiver:
Data format:
Floating Point Byte Order: ABCD
Termination:
Biasing:
HART® (Highway Addressable Remote Transducer) Protocol Information:
HART Version:
Device Description Files: DD files allow the user's handheld HART communicator to fully
Connections:
Protocol
Address (Modbus)
to position curser at Communication, and
to enter. When settings are completed press
2-wire, half-duplex
8 Data Bits
Jumper JP1 selectable 120Ω resistor.
TB1 & TB2 = OFF, TB2 & TB3 = ON
None
7.0
configure the GF UD2100. DD files for the Emerson 475
Communicator are provided. The files are included in the USB
drive provided with your GF UD2100 meter. You may also
request the files from GF via info.ps@georgfischer.com.
Warning: The GF UD2100 and associated DDs are pending
certification from the Fieldcomm Group.
HART Protocol uses a digital signal superimposed on the
4-20mA output. When the 4-20mA output of the GF UD2100 is
connected with a load resistor (230Ω to 600Ω), the HART
communicator can be connected on the loop in order to
communicate.
Choose MODBUS or HART.
Device address for the GF UD2100. Valid range: 001-247
(Default:
001). This number should be unique across the bus.
Press
or
store the setting.
to enter. Use
to scroll,