Herunterladen Inhalt Inhalt Diese Seite drucken

HBM FIT5A Bedienungsanleitung Seite 12

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Mechanical installation
application can be mounted on top. The screws and tightening torques to be
used are given in the table below:
Mounting
On mounting plate
For load application
1)
Recommended value for the specified property class. Please comply with the screw
manufacturer's instructions with regard to screw dimensions.
max. 70.5
∅20
Device socket 1
Fig. 4.1
Load application and installation
Important
The surface flatness of the components mounted on the load cell must be bet­
ter than 0.02 mm.
10
Thread
M6
M6
Platform
max. 30.5
Mounting base
(flatness v0.02 mm on
the mounting surface)
Device socket 2
(extended version E only)
A3916-2.0 HBM: public
Min. property class
10.9
10.9
Direction of loading
center of platform
Tightening torque
10 N⋅m
10 N⋅m
Adapter
(flatness v0.02 mm on
the mounting surface)
Alternative load
application
FIT®5A
1)